Features

Chemical identification and concentration in under 10 minutes VOC Detection in Air, Water or Soil Included Software Highly Portable Micro-GC Technology makes a FROG smaller than a laptop computer No specialty gases required — never runs out of carrier gas in the field Easy to program GC temperature profile Simple calibration process Lithium Ion Battery

By |2020-03-30T22:52:59+00:00December 23rd, 2019||Comments Off on Features

Authentication of the FROG – Air, Water, and Soil Analyses

Analysis of Solids CERTIFIED REFERENCE SOIL BY METHANOL EXTRACTION Analyte FROG-4000 (ug/kg) Certified Concentration (ug/kg) % Rec MTBE 8900 9450 94% Benzene 20400 20200 101% Toluene 49300 51400 96% Ethylbenzene 43700 40100 109% p/m-Xylene 28900 30400 95% o-Xylene 36300 33600 108% Air Analysis SIDE

By |2020-03-21T14:13:39+00:00December 23rd, 2019||Comments Off on Authentication of the FROG – Air, Water, and Soil Analyses

Authentication of the FROG – Independent Third Party Evaluation

Independent 3rd Party Evaluation BTEX IN WATER Analyte Result (ug/L) Assigned Value (ug/L) Acceptance Window Z Evaluation Benzene 8.1 6.97 5.26 to 8.67 1.99 Acceptable Ethylbenzene 8.0 9.33 0 to 32.7 -0.17 Acceptable Methyl tert-butyl ether (MTBE) 26.3 21.9 13.1 to 30.7 1.51 Acceptable Toluene 21.5 20.3 15.7 to 24.8

By |2020-03-21T14:02:50+00:00December 23rd, 2019||Comments Off on Authentication of the FROG – Independent Third Party Evaluation

Applications

AIR The FROG portable gas chromatograph can detect volatile organic compounds (VOCs) in air, delivering lab-quality results in less than 10 minutes. Read More Details About Air Applications WATER The FROG can detect volatile organic compounds in water as well. Water samples can be run onsite to

By |2020-04-04T16:28:34+00:00December 16th, 2019||Comments Off on Applications

Where to Use the FROG

IN THE FIELD The FROG Portable GC is rugged enough to be used at the job site while attached to a computer or stand alone with results appearing on the display. IN A LABORATORY Use the FROG Portable GC to save bench space and get results faster

By |2020-03-21T13:36:37+00:00December 9th, 2019||Comments Off on Where to Use the FROG

How to Use the FROG

WITH A COMPUTER You can initiate an analysis on the FROG portable GC using a computer. Attaching the FROG to a computer allows data to be studied in real-time with the Ellvin™ chromatography software. ON ITS OWN You can use the FROG portable GC on its own and

By |2020-09-22T19:28:43+00:00December 5th, 2019||Comments Off on How to Use the FROG

Specifications

Chemical Detection Range Vinyl Chloride (MW 62.5) 2-Methyl Napthalene (MW 142.2) Boiling Point @ 1 atm -13.3 °C 244.7 °C Vapor Pressure @ 25 °C 2980 mmHg 0.067 mmHg Instrument Specifications PID Lamp 10.6 eV energy GC Column 5.2 meter Weight 4.8 lbs. Dimensions 11 x 7.75 x 12 inches Battery Life Nine

By |2020-03-21T13:25:24+00:00December 3rd, 2019||Comments Off on Specifications

Chemicals Commonly Detected by the FROG

Benzene Toluene Ethylbenzene Xylenes Vinyl Chloride trans-1,2-Dichloroethene cis-1,2-Dichloroethene Trichloroethene (TCE) Tetrachloroethene (PCE) Naphthalene The FROG™ can also detect many other solvents

By |2020-03-27T14:19:56+00:00December 3rd, 2019||Comments Off on Chemicals Commonly Detected by the FROG