How long can I run the FROG-5000™ using battery power?
The FROG-5000™ has a Lithium Ion battery which lasts for 9 hours when fully charged.
The FROG-5000™ has a Lithium Ion battery which lasts for 9 hours when fully charged.
Size of box: 20"x14"x16" Weight: 24lbs ECCN: 1A995 Harmony Code: 9027.20.2000 (Gas Chromatograph) Country of Origin: USA Hazardous Goods: NO
Routine Maintenance on the FROGs include exchanging the scrubber material. Calibration is also required every 3 weeks for proper analysis with the FROG. Defiant also offers an annual service contract available for purchase.
When the FROG is calibrated at a different location than the one it is being used at adjusting retention times is necessary, due to differences in barometric pressure and/or altitude. You can adjust retention times by running the retention time check sample and following the instruction sheet provided with your calibrated Frog. It is important
The FROG uses air as the carrier gas and uses scrubber material (activated carbon and molecular sieve) to clean the air. The frequency at which these need to be replaced varies with how often the Frog is used and the cleanliness of the environment it is used in. For daily FROG use we recommend replacing